132
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
*SRE (Service Request Enable)
Command Syntax
*SRE <mask>
<mask> ::= integer with values defined in the following table.
The *SRE command sets the bits in the Service Request Enable Register.
The Service Request Enable Register contains a mask value for the bits to
be enabled in the Status Byte Register. A one in the Service Request
Enable Register enables the corresponding bit in the Status Byte Register.
A zero disables the bit.
Table 43 Service Request Enable Register (SRE)
Bit
Name
Description
When Set (1 = High = True), Enables:
7
OPER
Operation Status Register
Interrupts when enabled conditions in the
Operation Status Register (OPER) occur.
6
---
---
(Not used.)
From
Operation
Status
Registers
1
0
2
3
4
5
6
7
+
OPER
TRG
MAV
ESB
RQS/
MSS
TRG
Reg
TER?
Trigger Event Register
*SRE
*SRE?
*STB?
Status Byte Register
Service Request Enable
(Mask) Register
Output
Queue
SRQ
From
Standard
Event
Status
Registers
USR
MSG
Service Request
OR
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...