Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
165
:OPERegister:CONDition (Operation Status Condition Register)
Query Syntax
:OPERegister:CONDition?
The :OPERegister:CONDition? query returns the integer value contained in
the Operation Status Condition Register.
Wait
Trig
Run
Arm
Reg
AER?
Run bit set if oscilloscope not stopped
:OPEReration:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
From Overload
Event Registers
To OPER bit in
Status Byte
Register
OVLR
+
OR
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
Wait
Trig
Run
:OPERation[:EVENt]?
Operation Status Event Register
OVLR
HWE
11
5
3
From Hardware
Event Registers
HWE
12
From Mask Test
Event Registers
MTE
MTE
9
Table 52 Operation Status Condition Register
Bit
Name
Description
When Set (1 = High = True), Indicates:
15-13
---
---
(Not used.)
12
HWE
Hardware Event
A hardware event has occurred..
11
OVLR
Overload
A 50
Ω
input overload has occurred.
10
---
---
(Not used.)
9
MTE
Mask Test Event
A mask test event has occurred.
8-6
---
---
(Not used.)
5
Wait
Trig
Wait Trig
The trigger is armed (set by the Trigger Armed
Event Register (TER)).
4
---
---
(Not used.)
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...