:ACQuire:SEGMented:COUNt
Command Syntax
:ACQuire:SEGMented:COUNt <count>
<count> ::= an integer from 2 to 2000 (w/8M memory) in NR1 format
The :ACQuire:SEGMented:COUNt command sets the number of memory
segments to acquire.
The segmented memory acquisition mode is enabled with the
:ACQuire:MODE command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments in the
current acquisition is returned by the :WAVeform:SEGMented:COUNt?
query.
The maximum number of segments may be limited by the memory depth
of your oscilloscope. For example, an oscilloscope with 1M memory allows
a maximum of 250 segments.
Query Syntax
:ACQuire:SEGMented:COUNt?
The :ACQuire:SEGMented:COUNt? query returns the current count setting.
Return Format
<count><NL>
<count> ::= an integer from 2 to 2000 (w/8M memory) in NR1 format
See Also
•
•
•
•
•
•
•
"Introduction to :ACQuire Commands"
Example Code
•
N O T E
This command is available when the segmented memory option (Option SGM) is enabled.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...