230
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:CHANnel<n>:PROBe:SKEW
Command Syntax
:CHANnel<n>:PROBe:SKEW <skew value>
<skew value> ::= skew time in NR3 format
<skew value> ::= -100 ns to +100 ns
<n> ::= {1 | 2 | 3 | 4}
The :CHANnel<n>:PROBe:SKEW command sets the channel- to- channel
skew factor for the specified channel. Each analog channel can be adjusted
+ or - 100 ns for a total of 200 ns difference between channels. You can
use the oscilloscope's probe skew control to remove cable- delay errors
between channels.
Query Syntax
:CHANnel<n>:PROBe:SKEW?
The :CHANnel<n>:PROBe:SKEW? query returns the current probe skew
setting for the selected channel.
Return Format
<skew value><NL>
<skew value> ::= skew value in NR3 format
See Also
•