Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:MEASure:DEFine
Command Syntax
:MEASure:DEFine <meas_spec>
<meas_spec> ::= {DELay | THResholds}
The :MEASure:DEFine command sets up the definition for measurements
by specifying the delta time or threshold values. Changing these values
may affect the results of other measure commands. The table below
identifies which measurement results that can be affected by redefining
the DELay specification or the THResholds values. For example, changing
the THResholds definition from the default 10%, 50%, and 90% values may
change the returned measurement result.
:MEASure:DEFine
DELay Command
Syntax
:MEASure:DEFine DELay,<delay spec>
<delay spec> ::= <edge_spec1>,<edge_spec2>
<edge_spec1> ::= [<slope>]<occurrence>
<edge_spec2> ::= [<slope>]<occurrence>
<slope> ::= {+ | -}
<occurrence> ::= integer
MEASure Command
DELay
THResholds
DUTYcycle
x
DELay
x
x
FALLtime
x
FREQuency
x
NWIDth
x
OVERshoot
x
PERiod
x
PHASe
x
PREShoot
x
PWIDth
x
RISetime
x
VAVerage
x
VRMS
x
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...