Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
321
:MEASure:DUTYcycle
Command Syntax
:MEASure:DUTYcycle [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion | MATH}
<digital channels> ::= DIGital0,..,DIGital15 for the MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:DUTYcycle command installs a screen measurement and
starts a duty cycle measurement on the current :MEASure:SOURce. If the
optional source parameter is specified, the current source is modified.
Query Syntax
:MEASure:DUTYcycle? [<source>]
The :MEASure:DUTYcycle? query measures and outputs the duty cycle of
the signal specified by the :MEASure:SOURce command. The value
returned for the duty cycle is the ratio of the positive pulse width to the
period. The positive pulse width and the period of the specified signal are
measured, then the duty cycle is calculated with the following formula:
duty cycle = (+pulse width/period)*100
Return Format
<value><NL>
<value> ::= ratio of positive pulse width to period in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
Example Code
•
N O T E
The signal must be displayed to make the measurement. This command is not available if
the source is FFT (Fast Fourier Transform).
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...