350
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:MEASure:VMIN
Command Syntax
:MEASure:VMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VMIN command installs a screen measurement and starts a
minimum vertical value measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:VMIN? [<source>]
The :MEASure:VMIN? query measures and outputs the minimum vertical
value present on the selected waveform.
Return Format
<value><NL>
<value> ::= minimum vertical value of the selected waveform in
NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
•
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...