Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
355
:MEASure:VTOP
Command Syntax
:MEASure:VTOP [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VTOP command installs a screen measurement and starts a
waveform top value measurement.
Query Syntax
:MEASure:VTOP? [<source>]
The :MEASure:VTOP? query returns the vertical value at the top of the
waveform. The top value of the pulse is normally not the same as the
maximum value.
Return Format
<value><NL>
<value> ::= vertical value at the top of the waveform in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
•
N O T E
This query is not available if the source is FFT (Fast Fourier Transform).
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...