374
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:MTESt:ENABle
Command Syntax
:MTESt:ENABle <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:ENABle command enables or disables the mask test features.
•
ON — Enables the mask test features.
•
OFF — Disables the mask test features.
Query Syntax
:MTESt:ENABle?
The :MTESt:ENABle? query returns the current state of mask test features.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
•