Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
521
:TRIGger:GLITch:LEVel
Command Syntax
:TRIGger:GLITch:LEVel <level_argument>
<level_argument> ::= <level>[, <source>]
<level> ::= .75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= ±(external range setting) in NR3 format
for external triggers (DSO models)
<level> ::= ±8 V for digital channels (MSO models)
<source> ::= {CHANnel<n> | EXTernal} for DSO models
<source> ::= {CHANnel<n> | DIGital0,..,DIGital15} for MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :TRIGger:GLITch:LEVel command sets the trigger level voltage for the
active pulse width trigger.
Query Syntax
:TRIGger:GLITch:LEVel?
The :TRIGger:GLITch:LEVel? query returns the trigger level of the current
pulse width trigger mode. If all channels are off, the query returns
"NONE."
Return Format
<level_argument><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...