Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
555
:TRIGger:LIN:ID
Command Syntax
:TRIGger:LIN:ID <value>
<value> ::= 7-bit integer in decimal, <nondecimal>, or <string>
from 0-63 or 0x00-0x3f
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :TRIGger:LIN:ID command defines the LIN identifier searched for in
each CAN message when the LIN trigger mode is set to frame ID.
Setting the ID to a value of "- 1" results in "0xXX" which is equivalent to
all IDs.
Query Syntax
:TRIGger:LIN:ID?
The :TRIGger:LIN:ID? query returns the current LIN identifier setting.
Return Format
<value><NL>
<value> ::= integer in decimal
Errors
•
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
":TRIGger:LIN:SIGNal:DEFinition"
•
N O T E
This command is only valid when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.