:TRIGger:LIN:STANdard
Command Syntax
:TRIGger:LIN:STANdard <std>
<std> ::= {LIN13 | LIN20}
The :TRIGger:LIN:STANdard command sets the LIN standard in effect for
triggering and decoding to be LIN1.3 or LIN2.0.
Query Syntax
:TRIGger:LIN:STANdard?
The :TRIGger:LIN:STANdard? query returns the current LIN standard
setting.
Return Format
<std><NL>
<std> ::= {LIN13 | LIN20}
See Also
•
"Introduction to :TRIGger Commands"
•
•
":TRIGger:LIN:SIGNal:DEFinition"
•