Commands by Subsystem
:TRIGger:SPI:FRAMing
Command Syntax
:TRIGger:SPI:FRAMing <value>
<value> ::= {CHIPselect | NOTChipselect | TIMeout}
The :TRIGger:SPI:FRAMing command sets the SPI trigger framing value. If
TIMeout is selected, the timeout value is set by the
:TRIGger:SPI:CLOCk:TIMeout command.
Query Syntax
:TRIGger:SPI:FRAMing?
The :TRIGger:SPI:FRAMing? query returns the current SPI framing value.
Return Format
<value><NL>
<value> ::= {CHIPselect | NOTChipselect | TIMeout}
See Also
•