Commands by Subsystem
:TRIGger:UART:WIDTh
Command Syntax
:TRIGger:UART:WIDTh <width>
<width> ::= {5 | 6 | 7 | 8 | 9}
The :TRIGger:UART:WIDTh command determines the number of bits (5- 9)
for each message "byte" for the serial decoder and/or trigger when in
UART mode.
Query Syntax
:TRIGger:UART:WIDTh?
The :TRIGger:UART:WIDTh? query returns the current UART width setting.
Return Format
<width><NL>
<width> ::= {5 | 6 | 7 | 8 | 9}
See Also
•