Commands by Subsystem
5
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
629
:WAVeform:POINts
Command Syntax
:WAVeform:POINts <# points>
<# points> ::= {100 | 250 | 500 | 1000 | <points mode>}
if waveform points mode is NORMal
<# points> ::= {100 | 250 | 500 | 1000 | 2000 | 5000 | 10000 | 20000
| 50000 | 100000 | 200000 | 500000 | 1000000 | 2000000
| 4000000 | 8000000 | <points mode>}
if waveform points mode is MAXimum or RAW
<points mode> ::= {NORMal | MAXimum | RAW}
The :WAVeform:POINts command sets the number of waveform points to be
transferred with the :WAVeform:DATA? query. This value represents the
points contained in the waveform selected with the :WAVeform:SOURce
command.
For the analog or digital sources, the records that can be transferred
depend on the waveform points mode. The maximum number of points
returned for math (function) waveforms is determined by the NORMal
waveform points mode. See the :WAVeform:POINts:MODE command (see
) for more information.
Only data visible on the display will be returned.
When the :WAVeform:SOURce is the serial decode bus (SBUS), this
command is ignored, and all available serial decode bus data is returned.
Query Syntax
:WAVeform:POINts?
The :WAVeform:POINts query returns the number of waveform points to be
transferred when using the :WAVeform:DATA? query. Setting the points
mode will affect what data is transferred (see the :WAVeform:POINts:MODE
command (see
) for more information).
When the :WAVeform:SOURce is the serial decode bus (SBUS), this query
returns the number of messages that were decoded.
Return Format
<# points><NL>
<# points> ::= {100 | 250 | 500 | 1000 | <maximum # points>}
if waveform points mode is NORMal
<# points> ::= {100 | 250 | 500 | 1000 | 2000 | 5000 | 10000 | 20000
| 50000 | 100000 | 200000 | 500000 | 1000000 | 2000000
| 4000000 | 8000000 | <maximum # points>}
if waveform points mode is MAXimum or RAW
N O T E
The <points_mode> option is deprecated. Use the :WAVeform:POINts:MODE command
instead.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...