:CHANnel:THReshold
Command Syntax
:CHANnel:THReshold <channel group>, <threshold type> [, <value>]
<channel group> ::= {POD1 | POD2}
<threshold type> ::= {CMOS | ECL | TTL | USERdef}
<value> ::= voltage for USERdef in NR3 format [volt_type]
[volt_type] ::= {V | mV (-3) | uV (-6)}
The :CHANnel:THReshold command sets the threshold for a group of
channels. The threshold is either set to a predefined value or to a
user- defined value. For the predefined value, the voltage parameter is
ignored.
Query Syntax
:CHANnel:THReshold? <channel group>
The :CHANnel:THReshold? query returns the voltage and threshold text for
a specific group of channels.
Return Format
<threshold type> [, <value>]<NL>
<threshold type> ::= {CMOS | ECL | TTL | USERdef}
<value> ::= voltage for USERdef (float 32 NR3)
N O T E
The :CHANnel:THReshold command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :POD<n>:THReshold command (see
) or
:DIGital<n>:THReshold command (see
) for the InfiniiVision 7000 Series
oscilloscopes.
N O T E
•
CMOS = 2.5V
•
TTL = 1.5V
•
ECL = - 1.3V
•
USERdef ::= - 6.0V to 6.0V
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...