Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
131
To select trigger Coupling
1
Press the
Mode/Coupling
key.
2
Press the
Coupling
softkey, then select
DC, AC,
or
LF Reject
coupling.
•
DC
coupling allows DC and AC signals into the trigger
path.
•
AC
coupling places a 10 Hz (3.5 Hz in 100 MHz bandwidth
models) high-pass filter in the trigger path removing any
DC offset voltage from the trigger waveform. The high-pass
filter in the External Trigger input path is 3.5 Hz for all
models. Use AC coupling to get a stable edge trigger when
your waveform has a large DC offset.
•
LF
(low frequency)
Reject
coupling places a 50 kHz
high-pass filter in series with the trigger waveform. Low
frequency reject removes any unwanted low frequency
components from a trigger waveform, such as power line
frequencies, that can interfere with proper triggering. Use
this coupling to get a stable edge trigger when your
waveform has low frequency noise.
•
TV
coupling is normally grayed-out, but is automatically
selected when TV trigger is enabled in the Trigger More
menu.
Note that Trigger Coupling is independent of Channel Coupling.
To change Channel Coupling see
To select trigger Noise Rejection and HF rejection
1
Press the
Mode/Coupling
key.
2
Press the
Noise Rej
softkey to select noise reject or press the
HF Reject
softkey to select high frequency reject.
•
Noise Rej
adds additional hysteresis to the trigger circuitry.
When noise reject is on, the trigger circuitry is less
sensitive to noise but may require a greater amplitude
waveform to trigger the oscilloscope.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......