6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
To use FlexRay triggering
The N5432A FlexRay triggering and decode option, along with
the DECOMSYS::BUSDOCTOR 2 protocol analyzer and an MSO
communications cable (available from DECOMSYS), allow a
6000 Series mixed-signal oscilloscope to:
•
Control the BUSDOCTOR (without a PC).
•
Trigger on FlexRay bus frames, times, or errors and correlate
them with other signals in a device under test.
•
Display decoded FlexRay bus data.
Eight of the oscilloscope’s 16 digital channels are used for the
BUSDOCTOR connection.
Modes of BUSDOCTOR Control/Operation
The oscilloscope and BUSDOCTOR 2 can be used in one of three
ways:
•
Oscilloscope controls BUSDOCTOR, asynchronous mode
monitoring (LAN connection required).
•
Oscilloscope controls BUSDOCTOR, synchronous mode
monitoring (LAN connection required).
•
PC controls BUSDOCTOR.
When the oscilloscope controls BUSDOCTOR, the system is a
standalone, time-correlated measurement system (no external
PC required).
Oscilloscope Controls BUSDOCTOR, Asynchronous Mode
Monitoring
For asynchronous analysis, you can set the FlexRay channel and
baud rate from the oscilloscope’s BUSDOCTOR Menu.
N O T E
For information on displaying decoded FlexRay bus data, see
.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......