Triggering the Oscilloscope
4
6000 Series Oscilloscope User’s Guide
163
Data needs to be stable during the whole high clock cycle or
it will be interpreted as a start or stop condition (data
transitioning while the clock is high).
6
Return to the previous menu; then, press the
Trigger:
softkey
to select one of the following I
2
C trigger conditions:
•
Start Condition
— The oscilloscope triggers when SDA data
transitions from high to low while the SCL clock is high.
For triggering purposes (including frame triggers), a
restart is treated as a start condition.
•
Stop Condition
— The oscilloscope triggers when data (SDA)
transitions from low to high while the clock (SCL) is high.
•
Missing Acknowledge
— The oscilloscope triggers when SDA
data is high during any Ack SCL clock bit.
•
Address with no Ack
— The oscilloscope triggers when the
acknowledge of the selected address field is false. The R/W
bit is ignored.
•
Restart
— The oscilloscope triggers when another start
condition occurs before a stop condition.
SDA
SCL
Address
R/
Ack
Data
Ack
Stop
Condition
Start
Condition
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......