Integrate
∫
dt (integrate) calculates the integral of the selected source.
You can use integrate to calculate the energy of a pulse in
volt-seconds or measure the area under a waveform.
∫
dt plots the integral of the source using the “Trapezoidal
Rule”. The equation is:
Where
I = integrated waveform
∆
t = point-to-point time difference
y = channel 1, 2, or function 1 + 2, 1 – 2, and 1 * 2 data
points
c
o
= arbitrary constant
i = data point index
In Delayed sweep horizontal mode, the
∫
dt function does not
display in the delayed portion of the display.
1
Press the
Math
key, press the
∫
dt
softkey, then press the
Source
,
Scale
, or
Offset
sofkeys if you want to change the
source, scaling, or offset for the integrate function.
•
Source
— selects the source for
∫
dt. The source can be
any analog channel, or math functions 1 + 2, 1 – 2,
and 1 * 2.
•
Scale
— lets you set your own vertical scale factors for
∫
dt expressed in unit-seconds/division, where units can
be V (Volts), A (Amps), or W (Watts). Units are set in
the channel
Probe
menu. Press the
Scale
softkey, then
turn the Entry knob to rescale
∫
dt.
n
I
n
= c
o
+
∆
t
∑
y
i
i=0
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......