Making Measurements
5
6000 Series Oscilloscope User’s Guide
235
measurement thresholds. Each analog channel and the
math waveform can be assigned unique threshold values.
3
Press the
Type
softkey to set the measurement threshold to
%
(percentage of Top and Base value) or to
Absolute
(absolute value.)
•
Percentage thresholds can be set from 5% to 95%.
•
The units for absolute threshold for each channel is set
in the channel probe menu.
•
When the
Source
is set to
Math
, the threshold
Type
can only
be set to
Percent
.
4
Press the
Lower
softkey, then turn the Entry knob to set
the lower measurement threshold value.
Increasing the lower value beyond the set middle value
will automatically increase the middle value to be more
than the lower value. The default lower threshold is 10%
or 800 mV.
If threshold
Type
is set to
%
, the lower threshold value can
be set from 5% to 93%.
5
Press the
Middle
softkey, then turn the Entry knob to set
the middle measurement threshold value.
Source
select
Threshold
Type
Lower
Threshold
Middle
Threshold
Upper
Threshold
Return to
previous menu
Absolute threshold hints
• Absolute thresholds are dependent on channel scaling, probe attenuation, and
probe units. Always set these values first before setting absolute thresholds.
• The minimum and maximum threshold values are limited to on-screen values.
• If any of the absolute threshold values are above or below the minimum or
maximum waveform values, the measurement may not be valid.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......