Making Measurements
5
6000 Series Oscilloscope User’s Guide
249
Overshoot
Overshoot is distortion that follows a major edge transition
expressed as a percentage of Amplitude. The X cursors show
which edge is being measured (edge closest to the trigger
reference point).
Figure 42
Automatic Overshoot measurement
Rising edge overshoot =
local Maximum - D Top
Amplitude
---------------------------------------------------------- X 100
Falling edge overshoot =
Base - D local Minimum
Amplitude
----------------------------------------------------------- X 100
local Maximum
Overshoot
Overshoot
local Minimum
Top
Base