256
6000 Series Oscilloscope User’s Guide
6
Displaying Data
Grid intensity
To adjust the grid (graticule) intensity, press
Display
&
Grid
and
use the Entry knob
to adjust the intensity.
Vectors (connect the dots)
The Agilent 6000 Series oscilloscopes are designed to operate
optimally with vectors on. This mode gives the most realistic
waveforms in most situations.
When enabled,
Vectors
draws a line between consecutive
waveform data points.
•
Vectors give an analog look to a digitized waveform.
•
Vectors allow you to see steep edges on waveforms, such as
square waves.
•
Vectors allow subtle detail of complex waveforms to be
viewed, much like an analog oscilloscope trace, even when
the detail is just a small number of pixels in size.
The oscilloscope switches vectors on whenever the acquisition
system stops. Digital channels on the mixed-signal oscilloscope
are not affected by the Display menu. They are always displayed
with peak detect and vectors on. They also only contain one
trigger worth of information.
Accumulating multiple acquisitions
Turning off infinite persistence does not clear the display. This allows you to
accumulate multiple acquisitions, stop acquisitions, and then compare future
acquisitions to the stored waveforms.
Clearing stored infinite persistence waveforms
In addition to clearing the display by pressing the Clear Display softkey, the
display is also cleared of previous acquisitions if you press the AutoScale key.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......