6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
set the threshold level (TTL, CMOS, ECL, or user defined) for
the selected digital channel group. The threshold value is
displayed in the upper-right corner of the display.
When
Ext
(External Trigger) is selected, its level can be adjusted
using the
Level
knob in the Trigger section of the front panel.
The trigger level is displayed in the upper right corner of the
display.
The line trigger level is not adjustable. This trigger is
synchronized with the power line supplied to the oscilloscope.
To use Pulse Width triggering
Pulse Width (glitch) triggering sets the oscilloscope to trigger on
a positive or negative pulse of a specified width. If you want to
trigger on a specific timeout value, use
Duration
trigger in the
Trigger
More
menu.
1
Press the
Pulse Width
key in the Trigger section of the front
panel to display the Pulse Width trigger menu.
2
Press the
Source
softkey (or rotate the Entry knob on
mixed-signal oscilloscopes) to select a channel source for the
trigger.
Source
Pulse
polarity
Qualifier
Qualifier time set
Trigger level
or threshold
Trigger
source
Polarity
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......