Making Measurements
5
6000 Series Oscilloscope User’s Guide
203
7
Calculate the phase difference using the formula below.
Figure 24
Cursors set to center of signal
Z-Axis Input in XY Display Mode (Blanking)
When you select the XY display mode, the time base is turned off. Channel 1 is
the X-axis input, channel 2 is the Y-axis input, and channel 4 (or the External
trigger on 2-channel models) is the Z-axis input. If you only want to see portions
of the Y versus X display, use the Z-axis input. Z-axis turns the trace on and off
(analog oscilloscopes called this Z-axis blanking because it turned the beam on
and off). When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V),
the trace is turned off.
θ
=
sin
second
∆
Y
first
∆
Y
--------------------------
1.031
1.688
------------- ;
θ
= 37.65 degrees of phase shift
=
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......