236
6000 Series Oscilloscope User’s Guide
5
Making Measurements
The middle value is bounded by the values set for lower
and upper thresholds. The default middle threshold is 50%
or 1.20 V.
•
If threshold
Type
is set to
%
, the middle threshold value
can be set from 6% to 94%.
6
Press the
Upper
softkey, then turn the Entry knob to set
the upper measurement threshold value.
Decreasing the upper value below the set middle value
will automatically decrease the middle value to be less
than the upper value. The default upper threshold is 90%
or 1.50 V.
•
If threshold
Type
is set to
%
, the upper threshold value
can be set from 7% to 95%.
Time Measurements
The default lower, middle, and upper measurement
thresholds are 10%, 50%, and 90% between Top and Base
values. See
“To set measurement thresholds”
other percentage threshold and absolute value threshold
settings.
The following figure shows time measurement points.
FFT measurements
When you make an X at Max or X at Min measurement on a math FFT function,
the resultant units will be in Hertz. No other time related automatic measurement
can be made on the FFT math function. Use the cursors to make other
measurements on FFT.
Period
Fall Time
Rise Time
+ Width
– Width
Upper
Middle
Lower
Thresholds
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......