Getting Started
1
6000 Series Oscilloscope User’s Guide
53
To calibrate the probes
The oscilloscope can accurately calibrate its analog oscilloscope
channels to certain active probes, such as InfiniiMax probes.
Other probes, such as the 10073C and 10074C passive probes,
do not require calibration. The Calibrate Probe softkey will be
grayed-out (displayed in faint text) when a connected probe
does not require calibration.
When you connect a probe that can be calibrated (such as an
InfiniiMax probe), the
Calibrate Probe
softkey in the channel’s
menu will become active. Connect the probe to the Probe Comp
terminal, and the probe ground to the Probe Comp ground
terminal. Press the
Calibrate Probe
softkey and follow the
instructions on the display.
Passive Probes Supported
The following passive probes can be used with the 6000 Series
oscilloscopes. Any combination of passive probes can be used.
N O T E
When calibrating a differential probe, connect the positive lead to the
Probe Comp terminal and the negative lead to the Probe Comp ground
terminal. You may need to connect an alligator clip to the ground lug to
allow a differential probe to span between the Probe Comp test point and
ground. A good ground connection ensures the most accurate probe
calibration.
Table 9
Passive Probes
Passive Probes
Quantity Supported
10070C
4
10073C
4
10074C
4
10076A
4
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......