Front-Panel Controls
2
6000 Series Oscilloscope User’s Guide
85
This menu lets you select additional probe parameters such
as attenuation factor and units of measurement for the
connected probe.
•
Probe Attenuation
•
Skew
— When measuring time intervals in the ns range,
small differences in cable length can affect the
measurement. Use
Skew
to remove cable-delay errors
between any two channels.
Probe the same point with both probes, then press
Skew
and turn the Entry knob to enter the amount of skew
between channels. Each analog channel can be adjusted
±100 ns in 10 ps increments for a total of 200 ns
difference.
The skew setting is reset to zero when you press
Save/Recall
&
Default Setup
.
•
Probe Units
— Press the
Units
softkey to select the proper
measurement unit for the connected probe. Select
Volts
for
a voltage probe and select
Amps
for a current probe.
Channel sensitivity, trigger level, measurement results,
and math functions will reflect the measurement units you
have selected.
•
Calibrate Probe
— See
Attenuation
factor
Measurement
units
Skew adjust
Return to
previous menu
Calibrate
probe
Probe calibration not needed, not available.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......