26
Installation Manual
1
Introduction
Operators Maintenance
Maintenance appropriate for the operator is:
•
Cabinet cleaning - clean the cabinet using a dry cloth only.
•
Cleaning Optical Connectors
•
Cleaning SMA Connectors
.
•
Ensure ventilating fan cover is clean.
Cleaning Optical Connectors
You should clean the optical connectors at regular intervals
using the following materials:
WA R N I N G
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
Description
Agilent Part Number
Compressed Air Can or Blow Brush
Isopropyl Alcohol
Lint-free swab
5080-5400
C A U T I O N
Do not insert any tool or object into the optical IN or OUT ports of the
instrument as damage to or contamination of the optical fiber may
result.