112
N9310A User’s Guide
6
Subsystem Command Reference
Sweep Subsystem
The sweep subsystem controls the settings and parameters
associated with an RF/LF/Amplitude sweep.
RF Sweep State
:SWEep:RF:STATe ON|OFF|1|0
:SWEep:RF:STATe?
This command initiates the signal generator to sweep in a
different manner, according to the sweep trigger or point
trigger settings:
•
If the sweep trigger and point trigger are both set to
IMMediate
, sending this command below initiates an
RF sweep immediately:
“:SWEep:RF:STATe ON”
•
If either the sweep trigger or point trigger is set to
KEY
,
sending the commands
“:SWEep:RF:STATe ON”
arms
an RF sweep.
Sending command
“:TRIGger IMMediate”
initiates
the RF sweep.
•
If either the sweep trigger or point trigger is set to EXT,
sending the commands
“:SWEep:RF:STATe ON”
arms
an RF sweep. No further command is needed to initiate
the RF sweep, but a proper external signal should be
input into the signal generator.
The query returns a current RF sweep state.
NOTE
To make an RF/amplitude sweep available at the RF OUT connector, make
sure you enable RF output. See
Range
On (1), OFF (0)
*RST value
OFF
Returned data
1, 0
Front panel access
Sweep > Sweep Mode > RF/Off
Summary of Contents for Keysight N9310A
Page 34: ...28 N9310A User s Guide 2 Getting Started ...
Page 58: ...52 N9310A User s Guide 3 Using Functions ...
Page 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Page 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Page 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Page 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...