22
N9310A User’s Guide
2
Getting Started
Turn On the Signal Generator
1
Toggle the line switch
to on the rear panel to On state.
The orange standby LED will light and the signal generator
has connected with the AC line power.
2
Press the standby switch
on the front panel. The green
Switch On LED will light and the signal generator boots up.
Self- initialization takes about 30
seconds, including self- test.
If it detects an error, it reports an error message. The signal
generator then defaults to a maximum frequency of 3 GHz
and a minimum amplitude of –127 dBm, then the signal
generator is ready for your current use. After power on, let
the signal generator warm up for
45 minutes for
stabilization.
NOTE
The front panel switch is a standby switch only; it is not a power switch.
To disconnect the signal generator from the line power, turn off the power
switch on the rear panel.