Generating an Amplitude Sweep
Assume you are going to generate an amplitude sweep with
the following settings:
• amplitude ranges from –80 to –60 dBm, at a frequency of
1 GHz
• 9 sweep points and 50 ms dwell time at each point
Refer to the following steps to configure and generate an
amplitude sweep:
1
Press
hardkey
2
Press
>
1
>
GHz
3
Press
hardkey
4
Press
Step Sweep
softkey
5
Press
Ampl Start > –80 > dBm
6
Press
Ampl Stop > –60 > dBm
7
Press
#
Points > 9 > Enter
8
Press
More(1/2) > Step Dwell > 50 > ms
9
Press
hardkey to enable the RF output.
10
Press
Return > Sweep Mode > Ampl
to enable the amplitude
sweep immediately. A
SWEEP
annunciator displays for the