Key Reference
4
N9310A User’s Guide
55
AM Depth
Pressing this softkey enters the value of AM depth. The
active entry area displays the current value of AM depth.
The AM Depth ranges from 0 % to 100 %, with minimum
increment of 0.1 %.
•
Default value:
0.0 %
•
Key sequence:
> AM Depth
AM Source
Pressing this softkey reveals a menu of choices for amplitude
modulation sources. It allows you to select an internal
source or an external source from MOD IN connector, or
select internal and external simultaneously.
Pressing
INT
connects an internally calibrated signal to the
modulator. This internal AM source defaults to a sine wave
at a frequency of 1.0000 kHz.
Pressing
EXT
connects an externally calibrated signal to the
modulator. You need also set the
EXT Coupling
, see
on next page.
Pressing
INT+EXT
connects both an internally calibrated
signal and an externally calibrated signal to the modulator
simultaneously.
•
Default value:
INT
•
Key sequence:
> AM Source
AM Rate
Pressing this softkey modifies the internal modulation
frequency. The active entry area displays the current value
of AM rate. The allowed value range is 20 Hz to 80 kHz. The
minimum increment is 0.1 Hz. when external AM source is
selected, this softkey is disabled.
•
Default value:
1 kHz
•
Key sequence:
> AM Rate
AM
AM
AM
Summary of Contents for Keysight N9310A
Page 34: ...28 N9310A User s Guide 2 Getting Started ...
Page 58: ...52 N9310A User s Guide 3 Using Functions ...
Page 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Page 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Page 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Page 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...