72
N9310A User’s Guide
4
Key Reference
Pulse Period
Pressing this softkey modifies the pulse period of the
internal pulse signal. The active function area displays the
current pulse period value. The allowed value range is 200
μ
s to 2 s, with the minimum increment of 1
μ
s. The pulse
period applies only to the internal pulse signal. If an
external pulse source is selected, the
Pulse Period
softkey
is
disabled.
•
Default value:
200
μ
s
•
Key sequence:
>
Pulse Period
Pulse Width
Pressing this softkey modifies the pulse width for the
internal pulse signal. The active entry area displays the
current pulse width value. The allowed value range is 100
μ
s
to 1 s, with the minimum increment of 1
μ
s. If an external
pulse source is selected, the
Pulse Width
softkey is disabled.
•
Default value:
100
μ
s
•
Key sequence:
Pulse
>
Pulse Width
Pulse
Summary of Contents for Keysight N9310A
Page 34: ...28 N9310A User s Guide 2 Getting Started ...
Page 58: ...52 N9310A User s Guide 3 Using Functions ...
Page 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Page 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Page 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Page 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...