Subsystem Command Reference
6
N9310A User’s Guide
109
Amplitude Subsystem
The amplitude subsystem controls the settings and
parameters associated with amplitude.
Continuous Wave Output Amplitude
:AMPLitude:CW <val> <unit>
:AMPLitude:CW?
This command sets the amplitude of CW output.
The query returns the current amplitude of CW output.
Range and
unit
–
127 to +13 dBm, (+20 dBm settable)
–
80 to +60 dBmV, (+67 dBmV settable)
–
20 to +120 dB
μ
V, (+127 dB
μ
V settable)
0.0001 to 1000 mV, (2238.8 mV settable)
0.1 to 1 000 000
μ
V, (2238800
μ
V settable)
Minimum
increment
0.1 dB
(
When using dBm/dBmV/dB
μ
V
)
0.1 mV (When using mV)
0.1
μ
V (Ampl
999.9
μ
V)
100
μ
V (Ampl
1000
μ
V)
*RST value
–
127.0 dBm
Returned
data format
<val> <unit>
Front panel
access
≤
≥
Amplitude
Summary of Contents for Keysight N9310A
Page 34: ...28 N9310A User s Guide 2 Getting Started ...
Page 58: ...52 N9310A User s Guide 3 Using Functions ...
Page 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Page 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Page 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Page 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...