Subsystem Command Reference
6
N9310A User’s Guide
117
Amplitude Start
:SWEep:AMPLitude:STARt <val> <unit>
:SWEep:AMPLitude:STARt?
This command sets the start amplitude in an amplitude
sweep. The start amplitude should always be lower than the
stop amplitude in an amplitude sweep.
The query returns the current start amplitude.
Amplitude Stop
:SWEep:AMPLitude:STOP <val> <unit>
:SWEep:AMPLitude:STOP?
This command sets the stop amplitude in an amplitude
sweep. The stop amplitude should always be higher than the
start amplitude.
The query returns the current stop amplitude.
Range and unit
–
127 to +13 dBm
Valid unit
dBm
Minimum increment
0.1 dB
*RST value
–
127.0 dBm
Returned data format
<val> <unit>
Front panel access
Sweep > Step Sweep > Ampl Start
Range and unit
–
127 to +13 dBm
Valid unit
dBm
Minimum increment
0.1 dB
*RST value
13.0 dBm
Returned data format
<val> <unit>
Front panel access
Sweep > Step Sweep > Ampl Stop
Summary of Contents for Keysight N9310A
Page 34: ...28 N9310A User s Guide 2 Getting Started ...
Page 58: ...52 N9310A User s Guide 3 Using Functions ...
Page 94: ...88 N9310A User s Guide 5 Programming Fundamentals 5 Select Display a list ...
Page 106: ...100 N9310A User s Guide 5 Programming Fundamentals ...
Page 168: ...162 N9310A User s Guide 6 Subsystem Command Reference ...
Page 188: ... Agilent Technologies Inc 2006 Printed in China June 2006 N9310 90001 N9310 90001 ...