Subsystem Command Reference
6
N9310A User’s Guide
119
Sweep Repeat
:SWEep:REPeat SINGle|CONTinuous
:SWEep:REPeat?
This command toggles the sweep repeat between single and
continuous.
The query returns the current sweep repeat state.
Sweep Trigger
:SWEep:STRG IMMediate|KEY|EXT
:SWEep:STRG?
This command sets the sweep trigger mode to one of the
following choices:
•
IMMediate
- Selecting this mode immediately triggers
signal generator
to sweep from the first point to the final
point in a sweep sequence automatically.
•
KEY
- Selecting this mode arms a sweep. The armed
sweep will be initiated by sending the
“
TRIGger:IMMediate
” command.
•
EXT
- Selecting this mode enables an external trigger
source to initiate an armed sweep.
The query returns the current sweep trigger mode.
Range
SINGle/CONTinuous
*RST value
CONTinuous
Returned data
SING, CONT
Front panel access
Sweep > Sweep Repeat
Range
IMMediate, KEY, EXT
*RST value
IMMediate
Returned data
IMM, KEY, EXT
Front panel access
Sweep > Sweep Trigger