General Power Meter Functions
2
N1913/1914A EPM Series Power Meters User’s Guide
63
Setting Limits
The power meter can be configured to verify the current measurement in
any measurement line against predefined upper and/or lower limit values.
The range of values that can be set for the upper and lower limits and the
default values depends on the measurement units in the currently selected
measurement line (see
Table 2-3
Range of values for window limits
Procedure
Set the limits as follows:
1
Press
,
to display the
Measurement Setup
menus.
2
Use the
and
keys to highlight the
Limits:
setting field.
3
Press
to check the
Limits:
setting field.
4
Use the
key to highlight the
Minimum Limits:
value field.
5
Press
to display the
Minimum Limit
pop- up.
Window
Units
Maximum
Minimum
Default
Maximum
Default
Minimum
dB
+200 dB
–180 dB
60 dB
–120 dB
dBm
+230 dBm
–150 dBm
90 dBm
–90 dBm
%
999.9 X%
100.0 a%
100.0 M%
100.0 p%
W
100.000 XW
1.000 aW
1.000 MW
1.000 pW
N O T E
Ensure you have selected the channel you wish to set up.
Meas Select