General Power Meter Functions
2
N1913/1914A EPM Series Power Meters User’s Guide
51
Step Detection
To reduce the filter settling time after a significant step in the measured
power the filter can be set to re- initialize upon detection of a step
increase or decrease in the measured power. Step detection can be set in
both manual and automatic measurement average modes.
Procedure
Set step detection as follows:
1
Press
. On dual channel meters select the required channel.
2
Use the
and
keys to select the
Step Detect
setting field.
3
Press
to check the step detection to on or off.
4
Press
key to close the
Channel Setup
screen.