N4962A System Overview
Serial BERT 12.5 Gb/s User Guide
23
2.3
Connector Care
Inspect the connectors for the following:
•
Worn or damaged threads
•
Scratches to mating surface
•
Burrs and loose metal particles
•
Dust or foreign material in the space surrounding the center pin (type
K only)
•
Ensure that female contacts are straight and aligned
Clean the connectors as described in the following procedure. Cleaning
connectors with alcohol shall only be done with the instruments power cord
removed, and in a well-ventilated area. Allow all residual alcohol moisture to
evaporate, and the fumes to dissipate prior to energizing the instrument.
1.
Remove any dust or loose particles using a low-pressure air source.
2.
Moisten a lint-free swab with isopropyl alcohol. Do not saturate the
swab.
3.
Minimize the wicking of the alcohol into the connector structure.
4.
Clean the mating plane surfaces and threads.
5.
Allow alcohol to evaporate, and then use a low-pressure air source to
blow surfaces clean.
6.
Make sure no particles or residue remains.
7.
Inspect connector for damage.
Summary of Contents for N4962A
Page 1: ...Agilent N4962A Serial BERT 12 5 Gb s User Guide...
Page 6: ......
Page 16: ...Getting Started 16 Serial BERT 12 5 Gb s User Guide...
Page 24: ...N4962A System Overview 24 Serial BERT 12 5 Gb s User Guide...
Page 36: ...System Details and Performance Specifications 36 Serial BERT 12 5 Gb s User Guide...
Page 60: ...Operation 60 Serial BERT 12 5 Gb s User Guide...
Page 86: ...Remote GPIB Interface 86 Serial BERT 12 5 Gb s User Guide...
Page 88: ...Copyright Agilent Technologies 2012 Third edition May 2013 Printed in Germany...