Operation
Serial BERT 12.5 Gb/s User Guide
49
Table 15. N4962A front panel – display selection controls
Text
Type
Description
Default
Display
->
Scroll
UP and DOWN
buttons
and 6 lights
Buttons select the measurement or
configuration option to display on the
display panel:
•
UP selects the previous display item;
•
DOWN selects the next display item.
The selection process wraps around the
display options (pressing DOWN at the last
item selects the first item).
The display options are:
•
BER
Displays the current measured bit error rate
(BER), defined as the total number of bit
errors divided by the total number of bits
(bit counters detailed in Table 9).
If the error detector is on (
Receiver
->On
light is ON), the BER shown is a
measurement that is being calculated in
real time. If the detector is off, the BER
shown is from the previous session
(captured while the detector was
previously on).
•
ε time (sec)
Displays the elapsed time of the BER
measurement, in seconds. If the error
detector is on, the elapsed time will
increment in real time. If the error detector
is off, the elapsed time shown is from the
previous session.
•
Ampl (V)
Displays the single-ended PRBS output
amplitude, in volts. This voltage is per side,
multiply by 2 when the output is used
differentially.
Freq (GHz)
Summary of Contents for N4962A
Page 1: ...Agilent N4962A Serial BERT 12 5 Gb s User Guide...
Page 6: ......
Page 16: ...Getting Started 16 Serial BERT 12 5 Gb s User Guide...
Page 24: ...N4962A System Overview 24 Serial BERT 12 5 Gb s User Guide...
Page 36: ...System Details and Performance Specifications 36 Serial BERT 12 5 Gb s User Guide...
Page 60: ...Operation 60 Serial BERT 12 5 Gb s User Guide...
Page 86: ...Remote GPIB Interface 86 Serial BERT 12 5 Gb s User Guide...
Page 88: ...Copyright Agilent Technologies 2012 Third edition May 2013 Printed in Germany...