One-Button Measurements
5
N9320A Spectrum Analyzer
117
Res BW
Auto Man
Specifies the resolution bandwidth used in measuring the
offset pair. When set to
Auto
, the Res BW from the default
tables for
“Offset/Limits”
on page 114
are used. When set to
Man
, the allowed range is 1 kHz to 1MHz.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
Abs Start
Enables you to enter an absolute level limit at
Start Freq
for
the currently selected offset ranging from –200.00 to +50.00
dBm with 4 significant digits.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Abs Stop
Couple Man
Enters an absolute level limit at
Stop Freq
for the currently
selected offset ranging from –200.00 to +50.00 dBm with 4
significant digits, and to toggle this function between
Couple
and
Man
. If set to
Couple
, this is coupled to
Abs Start
to make
a flat limit line. If set to
Man
,
Abs Start
and
Abs Stop
you can
enter different values to make a sloped limit line.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Rel Start
Enters a relative level limit at
Start Freq
ranging from –150.00
to +50.00 dBc with 4 significant digits.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Rel Stop
Couple Man
Enters a relative level limit at
Stop Freq
ranging from –150.00
to +50.00 dBc with 4 significant digits, and to toggle this
function between
Couple
and
Man
. If set to
Couple
, this is
coupled to
Rel Start
to make a flat limit line. If set to
Man
,
Rel
Start
and
Rel Stop
you can enter different values to make a
sloped limit line.
Key access:
> More > Spectrum Emission Mask > Offset/Limits
> More
Fail Mask
Displays the menu to select one of the following logic keys
for fail conditions between the measurement results and the
test limits:
Absolute
,
Relative
,
Abs AND Rel
, and
Abs OR Rel
.
Key access:
> More> Spectrum Emission Mask > Offset/Limits
> More > More
Meas
Meas
Meas
Meas
Meas
Meas
Summary of Contents for N9320A
Page 1: ...Agilent N9320A Spectrum Analyzer User s Guide...
Page 4: ......
Page 18: ...10 N9320A Spectrum Analyzer 1 Overview...
Page 64: ...56 N9320A Spectrum Analyzer 3 Functions and Measurements...
Page 152: ...144 N9320A Spectrum Analyzer 8 Menu Maps MODE Menu Spectrum Analyzer Tracking Generator MODE...
Page 156: ...148 N9320A Spectrum Analyzer 8 Menu Maps SPAN Menu Span Full Span Zero Span Last Span SPAN...