5
Key Reference
106
N9342C/43C/44C User’s Guide
Delta Pair
Key access:
[MARKER] > {Mode}
Enters a mode that allows adjustment of both the
Ref
(start) and
Delta
(stop) markers independently.
Pressing
Delta Pair
, toggles between the reference
and delta markers. The start marker number is
indicated with a number and an R above the
marker (for example, 1R) and the delta marker is
indicated with a marker number. This mode is
useful in functions such as
Band Power
.
Span Pair
Key access:
[MARKER] > {Mode}
Enters a mode that allows adjustment of both the
ref and delta markers. Pressing
Span Pair
, toggles
between the span and center markers. The start
marker number is indicated with a number and an
R above the marker (for example, 1R) and the stop
marker is indicated with a marker number.
Adjusting the span changes the frequency
difference between the two markers while
maintaining the midpoint between the two markers
at a fixed frequency. Changing the center changes
the center point between the two markers while
maintaining the frequency difference. This mode is
useful in functions such as
Band Power
.
Off
Key access:
[MARKER] > {Mode}
NOTE
The delta marker function permits signal-to-noise
measurements provided the signal is a single spectral
component. Place a normal marker on the signal, press
Delta
,
place the delta marker in the noise, and activate Marker
Noise. The indicated amplitude indicates signal-to-noise/Hz.
NOTE
Ref and Delta markers maintain their displayed x-axis location,
but not their frequency values when you change a parameter
that redefines the x-axis scale. Reset these markers when
parameters such as Span or Center Freq are changed.
Summary of Contents for N9342C
Page 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Page 8: ......
Page 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Page 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Page 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...