Key Reference
5
N9342C/43C/44C User’s Guide
133
averages is greater than 12.5%, the digital filter
is cleared. The filter will then start storing new
measurement data. This feature shortens the
filter time required when the input power
changes substantially. Note that the result
appears to settle faster, although true settling to
the final value is unaffected.
•
Meas Interval
Sets the interval time between two
measurements. The default value is 100 ms.
Preset
Key access:
[MODE] > {Power Meter}
Sets the power meter softkey menu options to the
preset conditions. For example, the upper and
lower limits are set to their preset limit values.