Functions and Measurements
3
N9342C/43C/44C User’s Guide
53
Making a Reflection Calibration
The following procedure makes a reflection
calibration using a coupler or directional bridge to
measure the return loss of a filter. This example
uses a 370 MHz low- pass filter as the DUT. The
tracking generator (option TG7) is needed for this
measurement. For N9342C handheld spectrum
analyzer with option CA7 or CAU, please refer to
the
make a reflection measurement.
The calibration standard for reflection calibration
is usually a short circuit connected at the reference
plane (the point at which the DUT is connected). A
short circuit has a reflection coefficient of 1 (0 dB
return loss). It reflects all incident power and
provides a convenient 0 dB reference.
1
Connect the DUT to the directional bridge or
coupler as shown below. Terminate the
unconnected port of the DUT.
Figure 3-20
Reflection Measurement Short Calibration Test Setup
N9340A
HAN DHELD SPECTRU M ANALYZER
100 kH z - 3.0 GHz
PRESET
ENTER
FREQ
SPAN
AM PTD
BW/
SWP
SYS
M ODE
MEAS
TRACE
ESC/ CLR
2
D EF
3
GHI
1
ABC
5
M NO
4
JKL
6
PQR
8
VWX
7
STU
9
YZ_
0
SAV E
LIM IT
M ARKER
Coupled Port
Short
Circuit
DUT
or
2
Connect the tracking generator output of the
analyzer to the directional bridge or coupler.
NOTE
If possible, use a coupler or bridge with the correct test port
connector types for both calibrating and measuring. For the
best results, use the same adapter for the calibration and the
measurement. Terminate the second port of a two port device.
Summary of Contents for N9342C
Page 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Page 8: ......
Page 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Page 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Page 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...