3
Functions and Measurements
72
N9342C/43C/44C User’s Guide
Top N channel scanner measurement.
1
Connect the antenna to RF IN connector.
2
Turn on the channel scanner function.
Press
[Meas] > {Channel Scanner}
3
Press
[Meas] > {Scan Mode} > {Top N}
to set the
scan mode to TOP N channel mode. Input the
channel number you need to monitoring. Press
{Enter}
as a terminator.
4
Press
{Range Edit}
to access the submenu to edit
the scan range.
Toggle
{Edit Method}
to underline
Std
. It allows
you to set the range according to the embedded
communication standard, such as CDMA and
GSM which is used specifically in different
countries. Press
{Signal Std}
to select the
standard. Press
{Recall}
as a terminator. Press
{Begin Chn}
and
{End Chn}
to set the begin and end
channel for scanning. Press
{Apply}
to apply all
the range settings.
If you toggle
{Edit Method}
to underline
Custom
, it
allows you to set the customer- defined scan as
your need.
Press
{Start Freq}
to set the scan start frequency.
Press
{Freq Step Size}
to set the frequency interval
for adjacent channels.
Press
{Bandwidth}
to set the bandwidth of each
frequency channel which is used to calculate the
channel power.
Press
{Number of Chn}
to set the number of
frequency channels which will be scanned.
Press
{Apply}
as a terminator to apply all the
settings above.
5
Press
{Scan Start}
to initiate the channel scan
function.
6
Press
{Logging Start}
to start recording the scan
process which includes the channel power and
related channel bandwidth value to a CSV/KML
file. Press
{Logging Stop}
to end the logging. The
CSV/KML file will be saved immediately.
Summary of Contents for N9342C
Page 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Page 8: ......
Page 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Page 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Page 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...