Key Reference
5
N9342C/43C/44C User’s Guide
137
Trace
Accesses the trace keys to store and manipulate
trace information. Each trace comprises a series of
data points in which amplitude information is
stored. The analyzer updates the data for any
active trace with each successive sweep.
Trace
Key access:
[TRACE] > {Trace}
Selects the trace (1- 4) for current use.
Clear Write
Key access:
[TRACE] > {Clear Write}
Erases any previously stored trace data and
continuously updates the display during following
sweeps of the analyzer. This function is activated
for trace 1 at power on and factory
preset.
Max Hold
Key access:
[TRACE] > {Max Hold}
Maintains the displayed maximum level for each
trace point of the selected trace. A trace point is
updated if a new maximum level is detected in
successive sweeps.
NOTE
Changing the vertical scale (
Amplitude, Scale Type, Log or
Line
), or pressing
PRESET
, or turning averaging on (
TRACE,
Average (On)
) or switching widow in Zone Span restarts the
held trace.