Getting Started
24
N9342C/43C/44C User’s Guide
Adding an Option
Pressing
[System] > {More} > {Service} > {Add
Option}
brings up a dialog box for entering the option
license code. Use the numeric keypad to input the
option license code and then use the
[ENTER]
key as
a terminator. If the analyzer recognizes the option
license code, a message
“Option activated
successfully”
will appear in the status line.
Otherwise, a message
“Invalid option licence”
will
appear in the status line. Press
[System] > {System
Info} > {Installed Options}
to view the options.
Show Error
Pressing
[System] > {System Info} > {Error history}
accesses a list of the 30 most recent error
messages. The most recent error will appear at the
bottom of the list. If the error list is longer than 30
entries, the analyzer reports an error message
[–350, Query overflow]
. For more information,
refer to
Perform Calibration
The N934xC provides three manual calibration
function to calibrate the time base and amplitude.
The analyzer should warm up for 30 minutes
before calibration.
Time Base Calibration
Perform a time base calibration to guarantee the
frequency accuracy.
When the calibration function is triggered, the
current measurement is interrupted and a gauge
displays on the LCD. A message will display on the
LCD which indicates the calibration is finished,
and the interrupted measurement restarts.
Please refer to the operation procedures below:
1
Input a 10 MHz, 0 dBn signal to
EXT TRIG IN
.
2
Press
[System] > {More} > {Service} > {Calibration} >
{Time Base by Ext}
to initiate a calibration.
Summary of Contents for N9342C
Page 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Page 8: ......
Page 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Page 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Page 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...