129
Tandem Connection Monitoring
Tandem Connection Terminating Equipment (TCTE) Testing
Error Monitoring
Testing
Method: Use the OmniBER Transmitter to generate errors in the system under test
and verify on the Receiver that the correct error result is obtained.
13
Press the
key and select
.
14
Select the
folder.
15
Set the TEST FUNCTION field to
and select
.
16
Select an ERROR ADD TYPE and RATE, for this example we have chosen to
add B3 BIP errors at a rate of 1E-5, as shown below.
17
Press
to start the test. See "Test Timing " page 105 for advice on
setting test periods.
TRANSMIT
SDH
TEST FUNCTION
SDH
ERRORS & ALARMS
RUN/STOP
SONET_Usr.bk Page 129 Wednesday, December 13, 2000 11:11 AM