Performance Tests
3
Installation and Verification Manual
89
34
Set the oscilloscope to
Eye Measure
mode and setup to
measure the
Average Power
(in dBm) and
Extinction Ratio.
35
Check the Average Optical Power is between:
–5 dBm to +0 dBm (for 1310 nm)
–2 dBm to +3 dBm (for 1550 nm)
36
Check the Extinction Ratio is greater than
8.2 dB
and record
the measured value.
OC-48/STM-16 Eye Mask
N O T E
Ensure Vertical Calibration and Extinction Ratio Calibration have been
performed for the oscilloscope channel and vertical sensitivity range to
ensure measurement accuracy. Refer to oscilloscope information/Help if
necessary